Objective: The objective of the study was to develop analytical method for the estimation of tezacaftor and ivacaftor in bulk and its combination dosage form and to validate the method developed according to ICH guidelines. Materials and Methods: Chromatographic separation was achieved through Ascentis C18 150 × 4.6 mm, 5 m using 0.1% OPA: acetonitrile (70:30 v/v) mixture used as the mobile phase. The Water ACQUITY Model highperformance liquid chromatography system with photodiode array detector and EMPOWER version 2.0 software was monitored at detection wavelength 260 nm with flow rate of 1.0 ml/min and the method was validated as per ICH guidelines. Results and Discussion: By applying the proposed method, the retention times of tezacaftor and ivacaftor were found to be 3.0 and 3.6 min, respectively. %RSD of method precision for tezacaftor and ivacaftor was found to be 0.8 and 0.5, respectively. % recovery was obtained as 99.61% and 99.92% for tezacaftor and ivacaftor, respectively. Retention times were less so, the method developed was simple and economical and can be adopted in regular quality control analysis for selected drugs. Conclusion: This method was successfully applied for the determination of tezacaftor and ivacaftor in their pharmaceutical formulation and hence can be the routine analysis of these drugs in combined dosage form.