000 a
999 _c20245
_d20245
003 OSt
005 20231205103905.0
008 231205b xxu||||| |||| 00| 0 eng d
040 _aAIKTC-KRRC
_cAIKTC-KRRC
100 _922301
_a Adhao Vaibhav, S.
245 _aDevelopment and validation of stability indicating high performance liquid chromatography method for determination of leflunomide
250 _aVol.13(2), Apr-Jun
260 _aRaipur
_bAsian Pharma Press
_c2023
300 _a93-98p.
520 _aA new simple, specific, accurate and precise RP-HPLC method was developed for determination of Leflunomide. In the present study, stress testing of Leflunomide was carried out according to ICH guidelines Q1A (R2). Leflunomide was subjected to stress conditions of hydrolysis, oxidation, photolysis and neutral decomposition. Effective separation of drug and degradant was achieved was achieved on a Hypersil BDS C18 column (250mm × 4.6mm, 5.0µ particle size) under specific stress conditions using acetonitrile – 0.02M ammonium acetate buffer (60: 40, v/v) as a solvent system with a flow rate of 1.0mL/min. Quantification and linearity was achieved at 260nm over the concentration range of 5-30µg/mL for Leflunomide. The investigated method was validated as per guidelines.
650 0 _94639
_aPHARMACEUTICS
700 _912265
_aAmbhore, Jaya P.
773 0 _x2231-5667
_dRaipur A & V Publications
_tAsian journal of pharmaceutical analysis
856 _uhttps://ajpaonline.com/AbstractView.aspx?PID=2023-13-2-5
_yClick here
942 _2ddc
_cAR