Defrianto, Defrianto

External perspective of lung airflow model through diaphragm breathing sensor using fiber optic elastic belt - Vol.60(7), Jul - New Delhi CSIR 2022 - 561-566p.

Optical fiber-based detector technology is highly appreciated and developed in the field of medical physics. Through its fiber-optic wave pattern performance, this detector has great potential for monitoring difficult-to-calculate parameters such as airflow in the lungs. To realize this reality, a theoretical and experimental approach is needed in this research. The lung tissue model was formed using the Navier-Stokes equation using the finite element method by taking into account the continuity and momentum equations. While experimentally, single-mode fiber and fiber Bragg grating (FBG) was installed with a sinusoidal macro bending pattern as a strain sensor which was applied to the elastic belt and mounted on the diaphragm. The simulation model carried out depicts the velocity of air moving from the pulmonary duct to increase as it flows into smaller branches. While the experimental results show that the detected power parameter is a maximum of -0.16 dBm during inhalation and a minimum of -0.19 dBm during expiration. Due to the bending approach, the FBG sensor belt obtained the highest sensitivity at a sinusoidal bending diameter of 0.8 cm. Therefore, this is good news as a more accurate detector approach for medical purposes.


Humanities and Applied Sciences
Unique Visitors hit counter Total Page Views free counter
Implemented and Maintained by AIKTC-KRRC (Central Library).
For any Suggestions/Query Contact to library or Email: librarian@aiktc.ac.in | Ph:+91 22 27481247
Website/OPAC best viewed in Mozilla Browser in 1366X768 Resolution.

Powered by Koha